Diagnostic Accuracy of a Smartphone-based 12-lead ECG for Detection of Common Bradycardias, Tachycardias, and Ectopic Arrhythmias
No Thumbnail Available
Date
2026-02-24
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Galenos Publishing House
Abstract
Background and Aim: Electrocardiography (ECG) remains the cornerstone for diagnosing cardiac arrhythmias. The purpose of this study was
to assess the diagnostic performance of a smartphone-based ECG device (Spandan Ultra 12-lead) compared with a standard 12-lead ECG device
for detecting common cardiac arrhythmias (bradycardias, tachycardias, and ectopic arrhythmias), with a cardiologist as the reference standard.
Materials and Methods: The study was a prospective, cross-sectional, single-blind, observational, comparative diagnostic accuracy study
conducted in 321 patients aged ≥20 years who exhibited signs of arrhythmias. For analysis of diagnostic performance, sensitivity, specificity,
negative predictive value (NPV), positive predictive value (PPV), accuracy, F-score, positive and negative likelihood ratios [(positive likelihood
ratio (PLR) and negative likelihood ratio (NLR)], Matthews correlation coefficient (MCC), and the Farrington-Manning score were used to provide
a comprehensive evaluation.
Results: The mean age was 51.95±14.51 years; 50.47% were male. The smartphone-based ECG demonstrated higher sensitivity (86.54% vs.
77.80%), specificity (93.68% vs. 91.38%), PPV (72.58% vs. 64.61%), NPV (97.29% vs. 95.31%), accuracy (92.52% vs. 89.09%), F1 score (0.79 vs. 0.70),
PLR (13.7 vs. 9.05), MCC (0.75 vs. 0.64), area under the curve (0.851 vs. 0.846), and lower NLR (0.14 vs. 0.24) compared with the standard 12-lead
ECG. The Farrington-Manning non-inferiority test demonstrates that the smartphone-based ECG was non-inferior to the standard 12-lead ECG
on all validation parameters.
Conclusion: The Spandan Ultra 12-lead smartphone-based ECG is a reliable diagnostic tool for detecting common bradycardias, tachycardias,
and ectopic arrhythmias. Its simultaneous multichannel recording enables rapid and accurate rhythm assessment, demonstrating comparable
diagnostic performance to that of the standard 12-lead ECG and serving as a complementary diagnostic tool for detecting common arrhythmias.
Description
Keywords
Cardiac arrhythmias, 12-lead ECG, portable ECG, medical device
Citation
Mohan C, Gururani K, Rawat A, Singh Y, Chandola N, Agarwal D. Diagnostic Accuracy of a Smartphone-based 12-lead ECG for Detection of Common Bradycardias, Tachycardias, and Ectopic Arrhythmias. Int J Cardiovasc Acad. 2026 Jun 22;12(2):78-87. doi: 10.4274/ijca.2026.27928. Epub 2026 Feb 24.